Fresenius 6008 User Manual
The Fresenius 6008 CAREsystem improves hemodialysis efficiency and safety by reducing manual handling, featuring an all-in-one 6008 CAREset with pre-connected bloodlines, and automating tasks such as priming and rinsing. The system standardizes HighVolumeHDF and includes advanced monitoring tools, such as the Venous Access Monitor (VAM) and Online Clearance Monitor (OCM), designed for improved patient outcomes and reduced staff cognitive load. For more details, visit Fresenius Medical Care. 6008 CAREsystem - Fresenius Medical Care
| Error Code | Display Message | User Manual Solution | | :--- | :--- | :--- | | E-14 | "Blood Leak Detected" | Stop blood pump. Check dialyzer for rupture. Do not reuse dialyzer. | | E-22 | "Venous Pressure Out of Range" | Check for kinked lines or patient access stenosis. | | E-89 | "Conductivity Low" | Check concentrate caps. Prime the A/B lines to remove air. | | A-31 | "Bypass Active" | Check dialysate temp. Wait for normalization (max 15 mins). | fresenius 6008 user manual
Chapter 1: Safety and Indications for Use
- Warning: The machine must be connected to a hospital-grade grounded outlet. Use of unapproved power strips voids the warranty.
- Contraindications: The 6008 is not approved for hemodiafiltration (HDF) with post-dilution if blood flow is below 150 mL/min.
Disinfection Cycles:
- Thermal Disinfection (Daily): Runs at 85°C for 30 minutes. Use this after every patient.
- Chemical Disinfection (Weekly): Use citric acid or peracetic acid. Critical step: Rinse twice after chemical disinfection to remove residue.
- Heat Disinfection (Monthly): 93°C for 60 minutes to eradicate biofilms.
The 6008 user manual outlines several innovative technologies integrated into the system to enhance therapeutic outcomes: HighVolumeHDF: Warning: The machine must be connected to a
Safety Monitors: Features integrated monitors like the VAM (Venous Access Monitor) to detect external blood loss and the BVM (Blood Volume Monitor) to prevent intradialytic hypotensive episodes. Accessing the Manuals Advanced Fluid Management overview - Fresenius Medical Care Disinfection Cycles: